Call Now : +91 869 305 0520

Agasthya 2013 Series Syn Gas Analyzer BI 7200

Agasthya 2013 series BI7200 Syn Gas Analyzer is designed considering the flexibility required in the environments that characterize syn gas and bio-mass applications.

The analyzer uses high-stability infrared sensors for the simultaneous measurement of CO, CO2 and CH4. In addition, the analyzer can also be supplied with a non-consumable, long-life thermal conductivity cell for H2 that compensates for the interference effects of CO, CO2 and CH4. This ensures that H2 will always read correctly regardless of the background gas composition.

Paramagnetic sensor can be used for O2 analysis as an option. All sensors are temperature-compensated for maximum analytical stability. Auto calibration functionality allows easy calibration without user intervention.

Boiler usage in Oil Refinery
  • 32-Bit Contex-M3 ARM Processor
  • 5.7″ QVGA TFT Touch Screen Display
  • Touch-screen display for gas reading
  • Built-in sample pump or pressure regulator
  • Tried and tested technology with proven reliability
  • High range on each channel available (up to 100%)
  • H2 reading is compensated for the interference effect of the other gases measured
  • Automatic moisture removal system included
  • RS 485, supports MODBUS protocol
  • Isolated 4-20 mA analog output for each gas
  • Hi / Low gas, low flow and other alarms available
  • Cabinet purge system available
  • PTFE sample lines & sample components (where prudent)
  • Method of Detection
  • NDIR infrared sensor for CO, CO2, CH4
    Electrochemical sensor for O2 (Paramagnetic sensor as option)
    Thermal Conductivity sensor for H2
  • Ranges Available
    • O2
      0-25%, 0-50%, 0-100% Resolution: 0.10%
    • CO
      0-10%, 0-50%, 0-100%, Resolution: 0.10%
    • CO2
      0-10%, 0-50%, 0-100%, Resolution: 0.10%
    • CH4
      0-50%, 0-100%, Resolution: 0.10%
    • H2
      0-5% 0-50%, 0-100%, Resolution: 0.10%
  • Accuracy & Repeatability
  • ± 2% of reading for all gases
  • Drift
  • Less than 2% of full scale per month
  • Response Time (T-90)
  • 20-30 seconds
  • Ambient Temperature Range
  • 4 to 50°C
  • Power
  • 80 to 230 VAC 50 Hz
  • Output
  • Isolated 4-20 mA standard
    RS 485, supports MODBUS protocol
  • Relay Output
  • 2 Nos. of potential free contacts