Trace Oxygen Analyzer
Agasthya 2005 Series
BI 330 (% – Paramagnetic Technology)

Agasthya 2013 Series Trace Oxygen Analyzer BI 330 measures Traces of Oxygen in percent level in process gas. For O2 measurement Paramagnetic technology is used.

Easy to maintain and service from user point of view. LCD, LED & Touchscreen display available. Analog 4-20 mA output with 2 nos. potential free relay contact. It has 4 different mounting configurations Wall, Panel, Rack & Tabletop.

  • 32-Bit Cortex-M3 ARM Processor
  • Long life Paramagnetic O2 sensor
  • RS 485, MODBUS protocol
  • PTFE tubing and corrosion-resistant components for internal sample health
  • Flow meter for flow regulation
  • Method of Detection
  • Paramagnetic sensor for O2
  • Range Available
  • 0-30% scalable
  • Resolution
  • 0.01%
  • Accuracy & Repeatability
  • ±0.5% of FS
  • Drift
  • ±0.2% of full scale per month
  • Response Time (T-90)
  • ≤30 seconds
  • Display
  • LCD, LED
  • Power
  • 80 to 230 VAC, 50 Hz
  • Power Consumption
  • 8 Watts
  • Digital Output
  • RS485 with MODBUS Protocol
  • Analog Output
  • 4-20 mA (max load resistance upto 500 Ω)
  • Relay Output
  • 2 nos. of potential free contacts
  • Humidity
  • 5-95% RH
  • Ambient Temperature Range
  • 4 to 50°C
  • Flow Meter
  • 0.1-1 LPM
  • Enclosure
  • Rack
  • Sample Connection
  • 1/4 inch S.S. Connection
  • Area Classification
  • General Purpose
  • N2 Generator Plant
  • Gas Manufacturer
  • N2 Plant Manufacturer
  • Glove Box

    Related Products